Archive for the ‘Uncategorized’ Category

Statim 2000 Waste Bottle Over Filling

Thursday, July 15th, 2010

This was a simple, but unusual problem that was posed to me today

It seems as though the Statim 2000 Autoclave Waste Bottle was overflowing and the Statim 2000 could not complete the sterilization cycle.

At first glance, the client thought all of the water was being dumped into the waste bottle during the cycle
But there is not a direct path between the reservoir and the exhaust bottle so in reality, that is not possible.

To troubleshoot the problem, it is necessary to understand the path and function of the water leading from the reservoir to the exhaust on the Statim 2000 autoclave.

When a program is selected and the start button is pushed, water is taken from the reservoir and sent to the boiler through the Statim 2000 water pump. The boiler in turn converts the water to steam and the steam is then injected into the cassette.

As the steam expands, it pushes on the air inside the cassette. The exhaust then kicks on, removing the air from inside the cassette, forcing it into the exhaust bottle.

During this “purging” process, is when you hear the noise and notice the Statim 2000 exhaust tubing kicking wildly.

After all of the air has been removed, the exhaust is sealed and remains closed until either more air is detected, or the unit enters the vent stage. All of these process are implemented & controlled by the Statim 2000 PC board, based on the feedback the various components involved are sending to it.

After thinking about this process, it made it reasonable to assume that the Statim Waste Bottle was not being emptied periodically, and that is the reason it was full of water.

So, after explaining the cause of the problem, I asked the good Doctor to empty the waste bottle, refill it to the minimum line and try another cycle.

He did, and the Statim 2000 worked perfectly!

The moral of the story is to check the exhaust waste bottle from time to time on your Statim 2000 Autoclave and empty it when needed.

Wally
www.AllClaveParts.com
“Autoclave Repair Made Simple”

Share and Enjoy:
  • Digg
  • del.icio.us
  • Facebook
  • NewsVine
  • Reddit
  • StumbleUpon
  • Google Bookmarks
  • Yahoo! Buzz
  • Twitter
  • Technorati
  • Live
  • LinkedIn
  • MySpace

Statim 2000 Autoclave – Cycle Fault 16

Friday, December 18th, 2009


What It Means When Your Statim 2000 Autoclave Displays Cycle Fault 16… And How To Fix It

When your Statim 2000 Autoclave displays the “Cycle Fault 16″ error code, it is telling you that the steam generator (boiler) temperature went above 171&deg C during the heat up phase of the cycle.

This condition is generally caused from water not reaching the boiler.

The Statim 2000 Autoclave is designed so water is sent to the boiler from the water reservoir, by an electric water pump designed especially for the Statim 2000 Autoclave.

While Cycle Fault 16 on your Statim 2000 Autoclave could mean your water pump is now defective, it is most often caused simply from the water filters being clogged.

These filters are located on both sides of the water pump on the Statim 2000 autoclave. Cleaning the filters is not difficult and something you should be able to do on your own. This is how we do it……..



How To Clean The Statim 2000 Autoclave Water Filters



WARNING: Unplug Your Statim 2000 Autoclave Before Proceeding


To clean the Statim 2000 Autoclave Filters, you must first remove the fittings housing them.


Tools Required:

  1. Wrenches (3/8, 7/16, 9/16 and 5/8)
  2. (1) Pair Hemostats: To Clamp Water Line Tubing between Reservoir &amp Water Pump (or you can drain the Reservoir)
  3. (1) Phillips Screwdriver (To Remove Statim 2000 Autoclave Cabinet Housing)

Tools Recommended:

Parts Needed: None

Start by removing the housing cover.

  1. Remove the Statim 2000 Autoclave Cassette and set it aside
  2. Using the Phillips Screwdriver, remove the 4 screws from the sides (2 on each side) by turning them in a counter-clockwise direction
  3. Again, using the Phillips Screwdriver, remove the 3 screws from the rear of the cabinet (one on each side of the compressor and one on the opposite end)
  4. Set the screws aside in a safe place
  5. Lift up on the rear of the Statim 2000 Autoclave housing and slide it forward gently. Once cleared, you can pick it up and set it on it’s side (be careful, there are ribbon cables connected..this is where the Statim 2000 Extension Test Cable comes in handy

Start with the end of the Statim 2000 Autoclave water pump with the larger (gray) tubing connecting to the reservoir.

  1. Use hemostats to clamp the line to stop the water from flowing, and then pull the water line tube off of the pump
  2. Fit the 5/8 open-end wrench onto the flats of the white plastic fitting (Do Not Attempt To Turn The White Plastic Fitting) and the 9/16 onto the brass nut. Turn the nut counter-clockwise while holding the plastic nut in place.
  3. The Statim 2000 Autcolave screen is located inside the fitting. Do not attempt to remove the screen from the fitting. Remove the white o-ring and set it aside so you do not lose it. Set the fitting aside.

    Now go to the other end of the Statim water pump.

  4. Remove nut with the small diameter tubing from the elbow with a 3/8 open end wrench by turning it counter-clockwise.
  5. Place a 5/8 open end wrench on the nut under the bracket that the elbow screws into. Now place the 7/16 wrench on the elbow and turn it in a counter-clockwise direction.


    (Warning: Be Careful Not To Move, Disconnect, Hit Or Otherwise Damage The Silver Colored Tube. This Is The Statim 2000 Autoclave Thermocouple And Has Glass On The Inside. If it is damaged, you will have to send it to the factory for replacement)

    Tip: If the elbow will not turn, remove the larger wrench from the fitting and turn the elbow counter-clockwise. Once removed, then using the two wrenches, separate the elbow from the fitting. The screen is located on top of the elbow. Do not attempt to remove it. Remove the white o-ring and set it with the other one you just removed from the other fitting.

  6. Place the fittings into the ultrasonic cleaner and give it a chance to clean them thoroughly. Rinse well and blow out the excess water.

Once clean, reinstall the fittings and reconnect the tubing.

Put in a Statim 2000 Autoclave Cassette and run a cycle.

If Cycle Fault 16 does not return to the display, you have fixed the problem! Remove the Statim Autoclave Cassette, re-install the cabinet and put it back to work! If it does return, you will need to test the water pump

Note: If you had sent your Statim 2000 Autoclave off for repair of this problem , it would have cost you $1400 + shipping in both directions, and you would have had to wait a minimum of 30 days to get it back!
You just repaired your Statim 2000 Autoclave in a matter of minutes and it didn’t cost you a penny!

Wally

www.AllClaveParts.com

“Autoclave Repair Made Simple”

Share and Enjoy:
  • Digg
  • del.icio.us
  • Facebook
  • NewsVine
  • Reddit
  • StumbleUpon
  • Google Bookmarks
  • Yahoo! Buzz
  • Twitter
  • Technorati
  • Live
  • LinkedIn
  • MySpace

How To Find & Fix Tuttnauer Autoclave Leaks Around The Door

Wednesday, December 16th, 2009


Tuttnauer Leaking Around The Door

Is It The Door Gasket (Seal) Or The Door Bellows?

How To Isolate Leaks & Take Corrective Action on The Tuttnauer M, MK, and Valueclave Sterilizers.

If your Tuttnauer Autoclave is Leaking around the door, you will not be able to reach or maintain the proper temperature or pressure.


If not corrected, continued use of the Tuttnauer Autoclave in this condition puts additional stress on other components, and can lead to a host of other problems.

Worse yet, Sterilization Will Not Take Place

You can quickly determine if the problem is a faulty Door Gasket or the Door Bellows is a simple and straightforward procedure. This is something you can do on your own, No Autoclave Technician Needed.

Start by visually checking the Chamber Door Gasket for any steam leaks, hissing, or water bubbles at the Door Bellows.

If steam is leaking at the Tuttnauer door closing device, then rotate the Door Gasket 180° to see if the leak follows it.

If the leak follows the Door Gasket, then replace the Gasket. If the leak does not follow the Door Gasket, then replace the Door Bellows.

Tuttnauer recommends the Door Gasket be replaced every 12 months. When was the last time you replaced your Door Gasket?

Replacing the Door Bellows at the same time as the Door Gasket is considered to be a Tuttnauer Preventative Maintenance Routine Best Practice

Properly maintaining your Tuttnauer Autoclave can lead to many years of trouble free service and significantly increase its lifespan

Wally

www.AllClaveParts.com
“Tuttnauer Autoclave Repair Made Simple”

Share and Enjoy:
  • Digg
  • del.icio.us
  • Facebook
  • NewsVine
  • Reddit
  • StumbleUpon
  • Google Bookmarks
  • Yahoo! Buzz
  • Twitter
  • Technorati
  • Live
  • LinkedIn
  • MySpace

How To Test & Replace The Pelton Crane HP Autoclave Heating Element(s)

Thursday, October 22nd, 2009

If you are experiencing heat related problems with your Pelton Crane HP Autoclave, the odds are pretty good you have a faulty heating element.  Here is how to know for sure

The Pelton Crane HP originally had 2 heating elements, rated at 800 watts each. Testing them is simple and something you can do.

How To Test  The Original Pelton Crane  HP 800 Watt Heating Elements

This test consists of measuring the amount of resistance in the element. To do this, first, turn the autoclave off and unplug it from the wall

WARNING: CURRENT MUST NOT BE PRESENT WHEN MEASURING RESISTANCE.

You will need a volt/ohm meter to conduct this test.  The volt/ohm meter doesn’t have to be anything fancy or expensive.  You can get one for less than $20.00  from Walmart, Lowes, Home Depot, etc

Disconnect the wires from the Pelton HP autoclave heating elements so you don’t get a false reading.  Mark the wires so you know which wire went where, so when you complete the test, you can hook them back up correctly and quickly.

Set your volt/ohm meter on 200 ohms (200 NOT 200k). The ohms range is located within the Omega (Ω)(look for this symbol) section of your meter. It is also the area with “k” ranges (ie: 2000k,. 200k, 20K and then 2000, 200).

It doesn’t matter which color probe goes where.   Be sure to test both elements.

The range you are looking for is 16.2 – 19.8 ohms.  If  the readings on your elements are outside of this range, then they are faulty and need to be replaced.

The original 800 watt heating elements on the Pelton Crane HP Autoclave have been replaced with a single 1600 watt element.  So, even if just one is bad, you will need to remove and discard both of them. 

But don’t do it just yet.  Leave them connected until you have the new HP replacement heating element in hand.  This will greatly simplify the installation process and eliminate any confusion over the wiring of the new element

And, don’t be confused when connecting the wires to  the new element.  The old elements had 4 wires and the new element has only 2 terminals. Simply look at  it as though the second element didn’t exist. 

 The two wires on the left become a single wire on the left and the 2 wires on the right become a single wire on the right

 

The Pelton Crane HP Heating Element

Your Pelton Crane HP Autoclave is a true classic, and you have just added a few more years to it’s life !

 Wally
www.AllClaveParts.com
“Autoclave Repair Made Simple”

Share and Enjoy:
  • Digg
  • del.icio.us
  • Facebook
  • NewsVine
  • Reddit
  • StumbleUpon
  • Google Bookmarks
  • Yahoo! Buzz
  • Twitter
  • Technorati
  • Live
  • LinkedIn
  • MySpace

How To Clean The Tuttnauer EZ Series Water Outlet Strainer

Wednesday, October 21st, 2009

The Water Outlet Strainer on your Tuttnauer EZ series should be cleaned on a monthly basis.

Tuttnauer’s design engineers made cleaning the Water Outlet Strainer on you EZ Autoclave easy.  No need to remove the cabinet to gain access.  No tools or Service Technican required

Just a few simple steps:

Start by unplugging your Tuttnauer autoclave from the wall outlet and allowing the autoclave to cool down before proceeding (the Strainer Cap is located on the exhaust line and gets extremely hot). 

Make sure there is no water or pressure remaining in the Tuttnauer autoclave chamber. 

The Water Outlet Strainer is located on the back of the Tuttnauer autoclave.  There are 4 components to the water strainer:

  1. The Strainer Cap
  2. A Gasket
  3. The Strainer Element
  4. The Strainer Housing     

 tuttnauer_strainer_assembly

To remove the Strainer Element for cleaning, turn the strainer clamp counter-clockwise.  Remove the cap and the gasket.  Pull the Strainer Element out.

Rinse the Strainer Element under running water.  Use a brush if necessary (a tooth brush works very well).  After the Strainer element is clean, reinstall it, along with the gasket and Cap

 

Share and Enjoy:
  • Digg
  • del.icio.us
  • Facebook
  • NewsVine
  • Reddit
  • StumbleUpon
  • Google Bookmarks
  • Yahoo! Buzz
  • Twitter
  • Technorati
  • Live
  • LinkedIn
  • MySpace