Is AllClaveParts On It’s Way To Becoming A Movie Star?

Tuesday, December 21st, 2010

Well, They Are Going To Put Us In A Movie……(Really, it’s just our Autoclave Parts that are going to be in the movie, but that’s good enough for us!)

Announcing

Warner Brothers Studios Selected AllClaveParts To Provide Them With The High Quality Autoclave Parts They Needed For Their Upcoming Movie:

“Contagion”

Starring

Matt Damon & Gwyneth Paltrow

Jude Law, Kate Winslet and Laurence Fishburne

Scheduled For Release October 21, 2011

As described on MovieNewz.com “Contagion follows the rapid progress of a lethal airborne virus that kills within days. As the fast-moving epidemic grows, the worldwide medical community races to find a cure and control the panic that spreads faster than the virus itself…….”

Can You Guess What Part The Autoclave Plays?

We are very proud & thankful Warner Brothers chose AllClaveParts to furnish the High Quality Autoclave parts they needed for their production

And now I ask you, if the High Quality Autoclave Parts furnished by AllClaveParts was chosen to help fight a world-wide epidemic…..don’t you think they will be perfect for your environment?

AllClaveParts.Com

“Autoclave Repair Made Simple”

Chemiclave 5000 Pressure Rises, But Drops

Saturday, September 26th, 2009

When your Chemiclave 5000 Builds pressure then drops, it usually means there is a leak somewhere.  Here is what to look for:

a) Check the Door Seal for leaks.  The Door Gaskets (door seals) should be replaced on a regular basis as part of the Chemiclave 5000 Preventative Maintenance Program

b) Check for internal tubing fittings, pressure attachments and hose clamp leaks

c) Check for Pressure Relief Valve for leaks.  Over time, the Pressure Relief Valve’s Seals on the Chemiclave 5000 do breakdown  

d) Check the Metering Valve for Leaks.  The Chemiclave 5000 Metering Valve is a workhorse, but do develop leaks over time.

Wally
www.AllClaveParts.com
“Autoclave Repair Made Simple”

Share and Enjoy:
  • Digg
  • del.icio.us
  • Facebook
  • NewsVine
  • Reddit
  • StumbleUpon
  • Google Bookmarks
  • Yahoo! Buzz
  • Twitter
  • Technorati
  • Live
  • LinkedIn
  • MySpace

Chemiclave 5000 – Pressure Too Low

Friday, September 25th, 2009

Pressure Too Low –  Chemiclave 5000 Will Not Reach Vaporization (Green) Zone

a) Check that reservoir is filled with Vapo-Sterilization Solution

b) Be sure the sterilizer is not being overloaded in use. Extremely large loads of heavy metal may prolong heating and pressure build-up time in the Chemiclave 5000

c) Are you wrapping instruments in multiple wraps of linen or cloth or are you lining the tray with cloth towel.?

Heavy cloth insulation may prolong heating and pressure build up time in the Chemiclave 5000.

Chemically pure, hard surface paper bags and liners, such as

Harvey Indicator Bags (sterilization bags/pouches) and tray liners should be used in place of multiple wrapped or folded cloth

d) Check your Chemiclave 5000 for leaks

e) Check your Chemiclave 5000 for proper temperature

Share and Enjoy:
  • Digg
  • del.icio.us
  • Facebook
  • NewsVine
  • Reddit
  • StumbleUpon
  • Google Bookmarks
  • Yahoo! Buzz
  • Twitter
  • Technorati
  • Live
  • LinkedIn
  • MySpace

Chemiclave 5000 – Door Is Hard To Open

Monday, September 21st, 2009

Tension on the Chemiclave 5000 door is preset at the factory. Adjustment of door tension may be desirable after the Chemicals have been in use. Door should never be adjusted to permit pressure to bypass gasket at less than 40 PSI

If the door on your Chemiclave 5000 is difficult to open when chamber is not pressurized, relieve tension on door. To relieve tension, close door and turn 9/16″ hex bushing (E-61) counter-clockwise.

Bushing is located between bridge (E-42) and door (E-59). Use wrench on end of tray handle (E-100) or thin 9/16″ end wrench and adjust UNTIL LATCH CAN JUST BE OPENED WITHOUT SAFETY HANDLE BREAKING AT HINGE. DO NOT BEND SPRING

Wally
www.AllClaveParts.com
“Autoclave Repair Made Simple”

Share and Enjoy:
  • Digg
  • del.icio.us
  • Facebook
  • NewsVine
  • Reddit
  • StumbleUpon
  • Google Bookmarks
  • Yahoo! Buzz
  • Twitter
  • Technorati
  • Live
  • LinkedIn
  • MySpace

MDT Harvey Mediclave Autoclave Repair Tips

Thursday, September 10th, 2009

This is the forum to find autoclave repair tips & ask questions for all MDT Harvey Autoclaves